Part I Service Scenarios
Update 111
The L1-L5 indicators remain on: fully charged or >
80%.
The L1-L4 indicators remain on: 60% < remaining
capacity ≤ 80%.
The L1-L3 indicators remain on: 40% < remaining
capacity ≤ 60%.
The L1-L2 indicators remain on: 20% < remaining
capacity ≤ 40%
The L1 indicator remains on: 10% < remaining
capacity ≤ 20%
The L1 indicator blinks: remaining capacity ≤
10%
5. Baterry status indicator
Indicates the charge/discharge status of the U-Bank.
When the battery is charging:
Not fully charged: the indicator lights in orange.
Fully charged: the indicator lights in green.
When the battery is discharging:
The remaining capacity ≥ 20%: the indicator
lights in green.
The remaining capacity < 20%: the indicator blinks
in orange.
The remaining capacity ≤ 5%: the indicator
blinks in orange quickly.
When the U-Bank is charging or when you are
pressing the power button, the indicator blinks in
orange for 5s, and other indicators remain off. It
indicates that the U- Bank may have malfunction.
Please stop using the U- Bank and contact the
Mindray service engineers or your sales
representative.
6. Power button Used for checking the U-Bank power status.
7. Detach lever Used for detaching the U-Bank from the system.