Part I Service Scenarios
Maintenance 49
Use a dry soft cloth to wipe off the dust on the equipment enclosure (exposed part). Alternatively, use a
dry soft cloth immersed with a little wild soapsuds to wipe away remaining stains, and then leave it for
air drying.
2.1.3.8 Cleaning Peripherals
Clean actually configured peripherals and skip check items for those not configured.
Table 2-4 Peripheral cleaning list
No. Item Description
1 Color or black/white
video printer
Use a soft dry cloth to clean off the dust and stains on the
cover. Remove the cover to clean the inside of the printer. It is
necessary to abide by the procedures in the user manual to
clean the printer.
2 Graph/text printer Use a soft dry cloth to clean off the dust and stains on the
cover. Remove the cover to clean the inside of the printer. It is
necessary to abide by the procedures in the user manual to
clean the printer.
3 Footswitch Use a dry soft cloth immersed with a little wild soapsuds to
clean off the dust and stains on the keys and cables of the
footswitch.
4 Barcode scanner Use a dry soft cloth to clean the glass board of the scanner, and
then clean off the dust on the cables and the support. If special
cleaning is required, strictly follow the instructions in the
scanner user manual.
2.1.4 Disinfecting the Main Unit
Only the system equipped with trackpad supports disinfection.
WARNING:
Use only Mindray approved disinfectants and methods listed in this
section to disinfect the main unit. Warranty does not cover damage
caused by unapproved substances or methods.
Do not mix disinfectants, as hazardous gases may result.
We make no claims regarding the efficacy of the listed chemicals or
methods as a means for controlling infection. For the method to
control infection, consult your hospital’s infection control officer or
epidemiologist.
CAUTION:
Never immerse any part of the main unit in liquids or allow liquid to
enter the interior.
Any contact of disinfectants with connectors or metal parts may
cause corrosion.
Do not pour or spray any liquid directly on the main unit or permit
fluid to seep into connections or openings.