Trio™ Service Manual 0070-10-0591-01 5 - 5
Appendix System Alarm Prompts
LOOSE CUFF The NIBP cuff is not connected
correctly
Re-connect the NIBP cuff.
AIR LEAK The NIBP cuff is not connected
correctly or there are leaks in
the airway
Check the connection of each
part or replace cuff and hose.
If message does not clear
contact Customer Service.
OVER PRESSURE NIBP is in an overpressure
condition.
Check for occlusion in cuff
and hose. Measure again. If
message does not clear
contact Customer Service.
SIGNAL SATURATED Problem occurs when
measuring NIBP. The system
cannot perform measurement,
analysis or calculation.
Check the connection of the
cuff and hose. Check
patient's condition. Measure
again, if the failure persists,
contact the manufacturer for
repair.
TIME OUT Problem occurs when
measuring NIBP. The system
cannot perform measurement,
analysis or calculation.
Check the connection of the
cuff and hose. Check
patient's condition. Measure
again, if the failure persists,
contact the manufacturer for
repair.
CUFF TYPE ERR NIBP cuff may be wrong size Check if the patient size is set
correctly. Replace existing
cuff with correct size cuff. If
the failure persists, contact
the manufacturer for repair.
PNEUMATIC LEAK NIBP airway has leaks Check the connection of cuff
and hose or replace cuff and
hose. If the failure persists,
contact the manufacturer for
repair.
MEASURE FAIL Problem occurs when
measuring NIBP. The system
cannot perform measurement,
analysis or calculation.
Check the connection of the
cuff and hose. Check
patient's condition. Measure
again, if the failure persists,
contact the manufacturer for
repair.
NIBP SYSTEM FAILURE Problem occurs when
measuring NIBP. The system
cannot perform measurement,
analysis or calculation.
Check the connection of the
cuff and hose. Check
patient's condition. Measure
again, if the failure persists,
contact the manufacturer for
repair.
PROBLEM/MESSAGE REASON SOLUTION
* XX represents all the parameter modules in the system such as ECG, NIBP, SpO
2
, IBP module, etc.