EasyManuals Logo

Mindray Z60 User Manual

Mindray Z60
259 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #130 background imageLoading...
Page #130 background image
8-4 Measurement
8.4 Measurement Accuracy
Table 1 Error of 2D Images
Parameter
Value Range
Error
Distance Full screen
Within ±3%; or when the measured value is less
than 40 mm, the error is less than 1.5 mm.
Area (Trace) Full screen
Within ±7%; or when the measured value is less
than 16 cm
2
, the error is less than 1.2 cm
2
.
Area (ellipse,
circle)
Full screen
Within ±7%; or when the measured value is less
than 16 cm
2
, the error is less than 1.2 cm
2
.
Circumference
/
Within ±10%
Angle Full screen Within ±3%
Distance (iScape) Full screen
Linear probe: within ±5%
Other probe: within ±10%
Table 2 Volume Measurements
Parameter
Value
Range
Error
Volume Full screen
Within ±10 %; or when the measured value is less than 64 cm
3
,
the error is less than 6.4 cm
3
.
Table 3 Time/Motion Measurements
Parameter
Value Range
Error
Distance Full screen
Within ±3 %; or when the measured value is less
than 40mm, the error is less than 1.5mm.
Time Timeline Display Within ±2%.
Heart rate
Timeline Display
Within ±4%.
PW velocity
Min ≤10cm/s;
Max ≥200cm/s
(For probe L14-6P:
min≤10cm/s; Max≥
150cm/s)
≤5% (correction angle ≤60º)
CW velocity
Min ≤10cm/s;
Max ≥200cm/s
≤5% (correction angle ≤60º)
NOTE:
Within the selected field range, the measurement accuracy is ensured within the
range mentioned above. The accuracy specifications are performance in the worst
conditions, or based on the real test for the system, regardless of acoustic speed
error.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Z60 and is the answer not in the manual?

Mindray Z60 Specifications

General IconGeneral
Power Supply100-240V AC, 50/60Hz
ApplicationsAbdominal, Cardiac, Obstetrics/Gynecology, Vascular, Musculoskeletal, Small Parts, Pediatrics, Urology
Storage500GB hard disk
PortsEthernet, USB, VGA
Imaging ModesB/M/Color/Power/PW/CW