166
Item Function No. Number Contents
Display position of decimal
point.
71 01
0 : No decimal point, 1 : 1234.5, 2 : 123.45,
3 : 12.345, 4 : 1.2345
A/D sampling rate
71 02
0 : 20 times/s, 1 : 200 times/s
Condition of over display
71 03
0 : Bigger than |Weighing capacity +9D|
1 : Exceeding to weighing capacity ±110%
2 : Smaller than -20D and larger than weighing
capacity +9D
unit
71 05
0 : No unit, 1 : g, 2 : kg, 3 : t, 4 : lb, 5 : N, 6 : kN
Bridge power supply
71 07
0 : DC 10V, 1 : DC 5V, 2 : DC 2.5V
Operating condition of zero
set
71 10
0 : Operation when it is steady
1 : Operation by unconditional
Effective range of zero set
71 11
00 ~ 30 (Unit : ±1 % of weighing capacity)
Target of zero tracking
71 12
0 : Gross value and net value (for load display),
1 : Gross value (only at gross value display)
Date width of zero tracking
71 13
00 ~ 99 (Unit : 0.1 D)
Time width of zero tracking
71 14
00 ~ 99 (Unit : 0.1 s)
Operation of power on zero
71 15
0 : Invalid, 1 : Effective
Operating condition of tare
weight cancellation
71 16
0 : Operation when it is steady and 0<GROSS
≦WEIGHING CAPACITY
1 : Operate unconditionally and 0<GROSS
≦WEIGHING CAPACITY
2 : Operation when it is steady and 0<GROSS
≦| WEIGHING CAPACITY |
3 : Operate unconditionally and 0<GROSS
≦| WEIGHING CAPACITY |
Date stored
71 17
0 : RAM, 1 : EEPROM
Power on clear
71 18
000 ~ 111 (0 : Invalid, 1 : Effective)
10
1
digit : Data of tare weight cancellation
10
2
digit : Data of preset tare weight cancellation
10
3
digit : Data of zero set and zero tracking
Method of setting gravity
acceleration compensation
71 25
0 : Set the district number
1 : Set the numerical value of gravity acceleration
District number of using
place
71 26
01 ~ 16 (Unit : District)
District number of calibration
place
71 27
01 ~ 16 (Unit : District)
Gravity acceleration value
setting of using place
71 28
9.000 ~ 9.999 (Unit : m/s2)
Gravity acceleration value
setting of calibration place
71 29
9.000 ~ 9.999 (Unit : m/s2)
Range switch operation
71 40
0 : Range change OFF,
1 : Changeable two ranges
2 : Changeable three ranges
Setting of scale interval of the
second rang
71 41
0 : 1 scale interval, 1 : 2 scale interval,
2 : 5 scale interval, 3 : 10 scale interval,
4 : 20 scale interval, 5 : 50 scale interval
Boundary value setting in the
second range
71 42
000000 ~ 999999
Setting of scale interval of the
third rang
71 43
0 : 1 scale interval, 1 : 2 scale interval,
2 : 5 scale interval, 3 : 10 scale interval,
4 : 20 scale interval, 5 : 50 scale interval
Boundary value setting in the
third range
71 44
000000 ~ 999999
・ The command of writing is changing the upper commands [7*] to [8*].
・ When calibration lock is valid, only reading is possible as for the item of command [71].