62
5: Zero clear Return the condition of display before the zero set and zero
tracking when the zero set or the zero tracking executed
6: Tare clear Execute the tare weight cancellation
7: Accumulation clear Clear the accumulation value and times
8: Measurement comparison Move to Measurement comparison value setting mode
9: Forced batch finish Complete the weighing operation compulsorily
8-14. Setting of external control I/O
Set the external control I/O which I/O connecter is mounted at rear panel of this instrument.
8-14-1. Setting method of the external control input
The operation of an external control input is selected by setting function F-60 to F-68.
Setting range: 00 ~ 23
0: OFF No setting
1: ON/OFF key The same operation as the ON/OFF key
2: Setting key The same operation as the Setting key
3: Accumulation key The same operation as the Accumulation key
4: Preset tare key The same operation as the Preset tare key
5: Tare key The same operation as the Tare key
6: Gross/Net key The same operation as the Gross/Net key
7: Zero key The same operation as the Zero key
8: F key The same operation as the F key
9: Zero clear Return the condition of display before the zero set and zero
tracking when the zero set or the zero tracking executed
10: Tare clear Return the condition of display before the tare weight
cancellation when the tare weight cancellation is executed.
11: Batch start Input the signal to start batch
12: Discharge start Input the signal to discharge batch
13: Changeover the flow and discharge
Changeover the measuring mode flow and discharge
(ON: Discharge, OFF: Batch)
14: Clear the last accumulated data Clear the last accumulated data
15: Accumulation clear Clear the accumulation value and the accumulation times
16: Error cancellation Cancel the Error
17: Print Input the print signal
18: Hold Input the holding signal
19: Emergency stop Input the emergency signal
20: Manual free fall compensation Do the free fall compensation
21: Display of Net Display changes to Net of the load display.
22: Finish the weighing data compulsorily
Input the signal to finish the weighing data compulsorily
23: Finish the flow compulsorily Input the signal to finish the flow compulsorily
24: All brand accumulation clear Clear the accumulation value and accumulation times for all
brands
25: Brand 1 Input brand number using BCD code
26: Brand 2 Input brand number using BCD code
27: Brand 4 Input brand number using BCD code
Default is set to [OFF]
The following items are level input, and become effective at the input of the shortening of 50 ms or
more.
[13: Changeover the flow and discharge], [18: Hold], [21: Display of Net]
Other items are pulse input, and become effective once at the pulse input of over 50 ms or more.