VII
Caution
Please observe the following conditions strictly when this instrument suits the JIS standard.
If neglected, there is a possibility of not suiting the above-mentioned standard.
(3) Setting Functions
As for the details of the function of the value of C Function and Function,
please refer the paragraph 5-2 and 7-2.
CF-03 [Condition of over display] shall be applied the value of 2.
CF-08 [Net weight sign inversion] shall be applied the value of 0.
CF-11 [Effective range of zero set] shall be applied with less than the value of 02.
CF-13 [Data width of zero tracking] shall be applied with less than 4% of weighing capacity.
CF-20 [Set the stability detection in condition display while the set mode 2 or set of preset tare]
shall be applied the value of 0.
CF-21 [Set the unit display in sub-display section while the set mode 2 or set of preset tare] shall
be applied the value of 0.
F-01 [Setting of digital filter] shall be applied with more than the value of 016.
F-02 [Setting of analog filter] shall be applied with less than the set value of 01.
F-05 [Setting of stabilization filter] shall be applied with more than the value of 092.
F-06 [Data width of stabilization filter] shall be applied with less than the set value of 020.
F-07 [Time width of stabilization filter] shall be applied with more than the value of 05.
F-10 [Stability detection data width] shall be applied with the set value of more than 20.
F-11 [Stability detection time width] shall be applied with more than the value of 10.
F-15 [Operating condition of the preset tare weight cancellation] shall be applied the value of 0
or 1.
[OL] is appeared in the load date section at the over load. [-OL] is appeared in the load data section at
the condition of gross value is minus