6
CYCLES
TYPE OF LOAD
MAX KG
220 V - 240 V ~ 50Hz
220 V - 230 V ~ 60Hz
120 V ~ 60Hz
S17
S22
17
22
28
17
22
28
B 134°C
Universal
B 121°C
Universal
B 134°C
Prion (*)
Unwrapped porous
materials
-
-
1,00
1,25
1,50
0,60
0,75
1,00
Porous materials in
single package
-
-
0,75
1,00
1,25
0,50
0,60
0,75
Porous materials in
double package
-
-
0,60
0,75
0,90
0,40
0,50
0,60
Solid and hollow
materials in single
package
-
-
3,00
4,00
5,00
2,00
2,25
2,50
Solid materials and
cables not packed
-
-
6,00
7,50
9,00
4,00
4,50
5,00
Solid and hollow
instruments in
double package
-
-
1,50
2,00
2,50
1,00
1,25
1,50
S 134°C
Hollow
Unwrapped hollow
instruments
-
-
6,00
7,50
9,00
4,00
4,50
5,00
S 134°C Solid
Unwrapped solid
and hollow
instruments “B”
6,00
7,50
6,00
7,50
9,00
4,00
4,50
5,00
Solid and hollow
instruments “B” in
single package
3,00
4,00
3,00
4,00
5,00
2,00
2,25
2,50
S 121°C Solid
S 134 °C Prion (*)
Solid and hollow
instruments “B” in
single package
3,00
4,00
-
-
-
-
-
-
N 134°C Solid
N 121°C Solid
Unwrapped solid
and hollow
instruments “B”
6,00
7,50
-
-
-
-
-
-
(*) NOT AVAILABLE FOR US MARKET
NOTE
Definition of hollow loads in accordance with standard EN13060:2014
In this manual the term "hollow loads" refers both to the elements defined as "narrow
lumen" (paragraph 3.18 EN 13060:2014) and to the elements defined as "simple
hollow" (paragraph 3.30 EN 13060:2014).
The term "hollow load B" refers ONLY to the elements defined as "simple hollow"
(paragraph 3.30 EN 13060:2014).