Monitored Therapeutics, Inc. © Copyright 2020 13 P/N 45-50035 Rev. K, 2020-10-21
11.3.3. The battery of the GoSpiro must be replaced only by a factory trained
technician.
12. CALIBRATION
The GoSpiro is factory calibrated and should not require recalibration
between factory servicing every two years. If errors in spirometer
performance are suspected, the patient should be tested in a hospital or
physician office laboratory. If performance errors are confirmed, the GoSpiro
should be returned to Monitored Therapeutics for servicing and
recalibration.
CAUTION: ATS/ERS recommendations indicate that all spirometers used in
laboratories be calibrated or have the calibration verified daily.
13. CLEANING
The case of the GoSpiro may be cleaned using a damp cloth. The mouthport
adapter (if used by a single patient) should be cleaned after each use by
removing it from the GoSpiro (Figure 7). Soak the mouthport adapter in 1
teaspoon of dishwashing detergent (such as Palmolive) mixed in 3.5 cups of
warm water for ten minutes. Following soaking, the mouthport adapter must
be rinsed clean and allowed to air dry before replacing it on the GoSpiro.
Figure 7. Remove mouthpiece adapter and soak
Clinic or Physician Offices:
The case of the GoSpiro may be cleaned using a damp cloth or hospital
approved disinfectant wipe. The mouthport adapter or filter are single
patient use accessories and should be discarded after each use.