Monitored Therapeutics GoSpiro Spirometer - Compatible Filter Sources; Example Compatible Filter; Mouthport Adapter Accessory Details; Mouthport Adapter Functionality

Monitored Therapeutics GoSpiro Spirometer
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Monitored Therapeutics, Inc. © Copyright 2020 16 P/N 45-50035 Rev. K, 2020-10-21
15.2.3. Filters that fit the GoSpiro can obtained from A-M Systems, Inc. (a-
msytems.com). These include:
15.2.3.1. VBMAX30E (VBMax™ PFT Filter, E-Series, 30mm OD) Catalog
Number: 146300 (oval port)
15.3. Mouthport Adapter (Part Number 45-90061)
15.3.1. The mouthport adapter connects to the turbine assembly or mouthport
filter and has a stainless steel screen to protect the GoSpiro from
secretion you might exhale during your test. It is designed to protect
your GoSpiro.
15.3.2. The mouthport adapter does not filter out bacterial or viruses. It must
not be used by more than one person. It will not protect from disease
transmission between patients.
Please contact your distributor or www.monitoredrx.com for pricing
and purchasing options. Note that other mouthpieces that have
standard 27 mm fittings may also be used.
16. MAINTENANCE
The GoSpiro is designed to require very low maintenance. Please observe the
following precautions:
16.1. Replace the Turbine Assembly with a new one every 12 months as
preventive maintenance.
16.2. If the Turbine Assembly is exposed to dust or material in the GoSpiro, follow
the cleaning procedures in Section 13.
16.3. Clean the mouthport adapter after each use.
WARNING: If the GoSpiro is used for more than one patient without a
mouthport filter, the turbine assembly must be replaced or undergo high level
disinfection before use on another patient to prevent infection.
17. SERVICING
Routine maintenance consists of regular calibration checks every two years
and cleaning of the transducer. The GoSpiro must be returned to the factory
every 2 years for transducer inspection and accuracy check unless local
guidelines require a more frequent check.
Please contact info@monitoredRx.com or call 1.614.761.3555 if your GoSpiro
requires service, repair or if you need technical assistance. Before returning
your product to Monitored Therapeutics, first obtain a Returned Goods
Authorization (RGA) number. No product should be returned to Monitored