Echo/Vascular User Manual (15381-C) Page 33 of 46 DCR-00172
Troubleshooting Guide
No Actuator Function
Actuator(s) Not Running.
• Power cord not plugged all
the way into wall receptacle.
• Push power cord securely
into receptacle.
• Power outlet receptacle not
supplying 120 VAC power.
• Check power availability or
plug unit into another
receptacle.
• The power cord may be
separated from the control
box.
• Securely press power cord
into control box.
• Battery may be drained.
• Replace or charge battery.
• Actuator cord may be
unplugged.
• Push actuator cords securely
into actuator receptacle.
• Product was overloaded and
tripped internal fuse in
control box.
• Product motion is limited at
certain heights and angles.
• Return the surface to a level
position to restore full range
of motion
• The product will not move
when hand control buttons
are pressed.
• The hand control buttons
may be locked. If the green
LED on the hand control is lit
then press the unlock button
for 3 seconds to unlock the
hand control functions.
Squeaking noises during
operation.
• Actuator pins are not
sufficiently lubricated
• Apply WD40 or similar
lubricant to actuator pins.
WARNING
• Do not modify this equipment without authorization of the manufacturer.
• Removal of secured covers may increase the risk of electrical shock. Refer servicing to
qualified personnel.
• The battery should be periodically inspected for damage. Replace the battery if necessary.