Echo/Vascular User Manual (15381-C) Page 33 of 46 DCR-00172
Troubleshooting Guide
SYMPTOM
PROBABLE CAUSE
SUGGESTION
No Actuator Function
Actuator(s) Not Running.
Power cord not plugged all
the way into wall receptacle.
Push power cord securely
into receptacle.
Power outlet receptacle not
supplying 120 VAC power.
Check power availability or
plug unit into another
receptacle.
The power cord may be
separated from the control
box.
Securely press power cord
into control box.
Battery may be drained.
Replace or charge battery.
Actuator cord may be
unplugged.
Push actuator cords securely
into actuator receptacle.
Product was overloaded and
tripped internal fuse in
control box.
Replace control box.
Product motion is limited at
certain heights and angles.
Return the surface to a level
position to restore full range
of motion
The product will not move
when hand control buttons
are pressed.
The hand control buttons
may be locked. If the green
LED on the hand control is lit
then press the unlock button
for 3 seconds to unlock the
hand control functions.
Squeaking noises during
operation.
Actuator pins are not
sufficiently lubricated
Apply WD40 or similar
lubricant to actuator pins.
WARNING
Do not modify this equipment without authorization of the manufacturer.
Removal of secured covers may increase the risk of electrical shock. Refer servicing to
qualified personnel.
The battery should be periodically inspected for damage. Replace the battery if necessary.