Specifications
Echo/Vascular User Manual (15381-C) Page 36 of 46 DCR-00172
Environmental Conditions
ATTRIBUTE
Range for Use
Range for Storage and
Transport
Ambient Temperature
+0° to 40° C
-10° to +50° C
Relative Humidity
20% to 90% @ 30°C
not condensing
20% to 90% @ 30°C
not condensing
Atmospheric Pressure
860 to 1060 hPa
860 to 1060 hPa
Upholstery
ATTRIBUTE
SPECIFICATION
Foam
California Technical Bulletin 117
Vinyl
D.O.T. FAR 25.853 and FMVSS302
Port of NY / Boston F.D. Code
Power Requirements
ATTRIBUTE
SPECIFICATION
Electrical, Product
100240 VAC, 50/60 Hz, max 5 A
Electrical, Battery Charger
100240 VAC, 50/60 Hz
Battery Option
24 VDC, 2.9 Ah
Duty Cycle
10%, 2 min. on / 18 min. off
All electrical circuitry is isolated from chassis.
Grounding reliability can only be achieved when the equipment is connected to
an equivalent receptacle marked “Hospital Only” or “Hospital Grade.
The power cord is to be used for mains disconnection.
Attached power supply cord set is not allowed to be diverted to other
equipment.