2-2 Electrical Specifications (23 5 )
Uni
t
Ran
ge
In-ran
g
e Displa
y
Resolutio
n
40.00 Lux 0 to 40.00 Lux 0.01 Lux
Lux 400.0 Lux 36.0 to 400.0 Lux 0.1 Lux
4,000 Lux 360 to 4,000 Lux 1 Lux
40,000 Lux 3,600 to 40,000 Lux 10 Lux
400,000 Lux 10,000 to 400,000 Lux 100 Lux
4.000 Fc 0 to 3.720 Fc 0.001 Fc
Foot- 40.00 Fc 3.35 to 37.20 Fc 0.01 Fc
candle 400.0 Fc 33.5 to 372.0 Fc 0.1 Fc
( Ft-cd ) 4,000 Fc 335 to 3,720 Fc 1 Fc
40,000 Fc 930 to 37,200 Fc 10 Fc
Foot-candle = Ft-cd = F
c
Uni
t
Ran
ge
Accurac
40.00 Lux
Lux 400.0 Lux ( 3% rd
g
+ 0.5 % F.S. )
4,000 Lux
40,000 Lux
400,000 Lux < 100,000 Lux :
( 3% rd
g
+ 0.5 % F.S. )
100,000 Lux :
@ for reference onl
y
4.000 Fc
Foot- 40.00 Fc ( 3% rd
g
+ 0.5 % F.S. )
candle 400.0 Fc
( Fc ) 4,000 Fc
40,000 Fc < 9,300 Fc :
( 3% rd
g
+ 0.5 % F.S. )
9,300 Fc :
@ for reference onl
y
Note : Accuracy tested by a standard parallel light
tungsten lamp of 2856 temperature.