EasyManua.ls Logo

MSA Ultima - Alarm Status Screen; Sensor

MSA Ultima
30 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Sensor:
• Automatically recognized when connected to a transmitter
• When removed, causes a "Sensor Missing" fault; this can be
corrected by:
• Reconnecting a sensor to that position or
• Manually taking the sensor 'off-line' via:
• Ultima X Controller, by sending a sensor disable command
• ModBUS command write to a control register.
• A missing sensor or unused sensor position returns a gas value of
-99.9 in response to a ModBUS request for gas level value starting
at address base +207.
Alarm Status Screen
(FIGURE 2-3)
In the North American version of the Ultima X with X3 Technology, a
screen will appear when a combustible sensor crosses 50%, 60% or
90% of the full-scale range.
• This screen:
• will override all sensor data or fault screens and will clear when
all sensors are below 50% of scale.
• can be acknowledged by the operator; the:
• instrument will go back to normal display operation until the
sensor or any other sensor crosses 50%, 60% or 90% of
scale
• normal operation status screen will display again.
• can be acknowledged through the:
• Ultima X Controller
• Pushbutton or
• MODBUS interface.
NOTE: All IR inputs will be functional and appropriate display
screens will display for the received IR message.
2-4