30
[1]
The rotary switch position is only for specific models.
[2]
When
measuring current, for 10 A to 15 A, the measuring duration should not be over
2 minutes within 10 minutes, and in this 10 minutes, no other current should flow
through except within the measuring duration; for 15 A to 20 A, the measuring
duration should not be over 10 seconds within 15 minutes, and in this 15 minutes, no
other current should flow through except within the measuring duration.
[3]
When
measuring capacitance, for the 60.00mF range, the measuring duration should
be over 30 seconds.
[4]
When measuring frequency, the typical waveform is Square or Sine. The signal meets
the following conditions.
Frequency
Amplitude (rms)
1 Hz – 5 MHz
≥ 700 mV
[5]
When measuring duty cycle, the typical waveform is Square.
Characteristics
Instruction
Display
5999
Frequency Response (Hz)
(40 - 1000) Hz
Sample rate for digital data
3 times/second
Bluetooth
MO730026
A
Without
MP730026
√
Auto ranging
√
True RMS
√
Diodes Test
√
Sleep Mode
√
Continuity Test
√
Low battery indication
√(The " " is displayed when the battery is
under the proper operation range.)
Data Hold
√
Relative Measurement
√
LCD Backlight
√
Input Protection
√
Input Impedance
≥ 10 MΩ
Battery
9V battery (6F22)
LCD Size
58.5 mm * 41 mm
Weight (without package)
0.32 kg
Dimension
190 mm * 90 mm * 56 mm
Working temperature
0℃ to 40℃
Storage temperature
-10℃ to 60℃
Relative Humidity
≤ 80%
Altitude
Operating: 3,000 m
Non-operating: 15,000 m
Interval Period of Adjustment: One year is recommended for the calibration interval period