11.5 Capacitance Measurement
Range
Resolution Accuracy
MP60607 MP60608
40.00nF 60.00nF 0.01nF
±(4%+5)
400.0nF 600.0nF 0.1nF
4.000uF 6.000uF 0.001μF
40.00uF 60.00uF 0.01μF
400.0uF 600.0uF 0.1μF
4.000mF 6.000mF 0.001mF
±10%
40.00mF 60.00mF 0.01mF
Caution:
• Measured value = displayed value - open circuit value of the test leads (For capacitance ≤100nF, “REL” mode is recom-
mended, open circuit has residual reading). The guaranteed accuracy is 1%~100%.
• Overload protection: 600Vrms
11.6 Frequency Measurement
Range Resolution Accuracy
10Hz~10MHz 0.01Hz~0.01MHz ±(0.1%+4)
Caution:
Measurement sensitivity:
≤100kHz: 200mVrms≤ input range≤30Vrms;
>100kHz~1MHz: 600mVrms≤ input range≤30Vrms;
>1MHz~10MHz: 1Vrms≤ input range≤30Vrms.
Duty ratio is only applicable to ≤10kHz square wave measurement with a range of 1Vp-p:
If frequency≤1kHz, duty cycle will be 10.0%-95.0%;
If frequency>1kHz, duty cycle will be 30.0%-70.0%.
11.7 Temperature Measurement (MP60608 only)
Range Resolution Accuracy
-40°C~+40°C
1°C
±4°C
40°C~500°C ±(1.5%+5)
500°C~1000°C ±(2.0%+5)
-40°F~104°F
1°F
±6°F
104°F~932°F ±(2%+6)
932°F~1832°F ±(2.5%+4)
Caution:
The meter displays “OL” after startup, it is only suitable for K-type thermocouple (Nickel-Chromium ~ Nickel-Silicon temperature
sensor) and temperature measurements below 1000°C/1832°F.
The formula for Celsius to Fahrenheit is °F =1.8°C + 32.