The CURRENT TIME field will automatically synchronize
with the PC date and time setting
The MANUAL and INSTANT select buttons allow the user to
start datalogging immediately when the Setup window is
exited (INSTANT) or at a later time (MANUAL). Explained in
the LOGGING READINGS section
The LOGGER NAME field allows the user to select a unique
name for the logger
The SAMPLE POINTS field instructs the logger to take a
finite number of readings
The SAMPLE RATE pull-down menu instructs the logger to
log readings at a specific rate
The LED Flash Cycle can be set by the user (the longer the
time, the longer the battery life)
The ALARM SETTINGS portion of the SETUP window
allows the user to set HIGH and LOW temperature/RH limits
and disable the flashing of the Logger’s LEDs thereby
conserving battery energy
Click on the SETUP button to save changes. Press the
DEFAULT button to set the Logger to factory default condition.
Press the CANCEL button to abort the setup.
Notes: Any stored data will be permanently erased when Setup is
finished. To enable you to save this data before it is lost, click
Cancel and then you need to download data.
The battery will in all likelihood run out before logger will have
finished specified sample points. Always ensure that the
remaining charge in the battery is sufficient to last the complete
duration of your logging exercise. If in doubt, we recommend that
you always install a fresh battery before logging critical data.