NI PXI-4130 Calibration Procedure | © National Instruments | 39
1 20 mA 20 V 20 V 4 6.80 mA
5 9.0 mA
6 11.2 mA
7 13.4 mA
8 15.6 mA
9 17.8 mA
10 20.0 mA
-20 V 11 0.200 mA
12 2.40 mA
13 4.60 mA
14 6.80 mA
15 9.00 mA
16 11.2 mA
17 13.4 mA
18 15.6 mA
19 17.8 mA
20 20.0 mA
Table 20. NI 4130 Output Parameters and Test Limits for Current Programming
Accuracy Adjustment (Continued)
Channel
Current
Limit
Range
Voltag e
Level
Range
Voltag e
Level
Iteration Output