10 | ni.com | NI PXI-5650/5651/5652 Calibration Procedure
4. Use the power meter to measure the RF OUT power. Compare the measured RF OUT
power to the Test Limit listed in Table 4.
Table 4. High Power Level Accuracy Test Points and Limits
Device Frequency
Start Power
(dBm)
Test Limit (dB)
PXI-5650/5651/
5652
500 kHz 5 ±1.6
1 MHz 5
2 MHz 5
5 MHz 5
10 MHz 8 ±0.75
20 MHz 8
50 MHz 10
100 MHz 10
150 MHz 10
200 MHz 10
500 MHz 10
1 GHz 10
PXI-5651/5652 1.5 GHz 10 ±0.75
2 GHz 8
2.5 GHz 8
3 GHz 8
3.3 GHz 7
PXI-5652 3.5 GHz 7 ±1.0
4 GHz 5
4.5 GHz 5
5 GHz 0
5.5 GHz 0
6 GHz 0
6.6 GHz 0