3i Newborn Hearing Screener IFU 024913 RevD Revision Date 08/05/2020
48
Table 7 Patient Coupling Port
Phenomenon
Basic EMC Standard
Immunity Test Levels - Professional
Healthcare Facility Environment
Electrostatic Discharge
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Conducted Disturbances Induced by
RF Fields
IEC 61000-4-6
3 V
0,15 MHz 80 MHz
6 V in ISM bands between 0,15 MHz and 80
MHz
80% AM at 1 kHz
Table 8 Immunity Test Levels - Signal Input / Output Parts Port
Phenomenon
Basic EMC Standard
Immunity Test Levels - Professional
Healthcare Facility Environment
Electrostatic Discharge
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Electrical Fast Transients/Bursts
IEC 61000-4-4
± 1 kV
100 kHz repetition frequency
Surges Line-to-Ground
(Common Mode)
IEC 61000-4-5
± 2 kV
Conducted Disturbances Induced by
RF Fields
IEC 61000-4-6
3 V
0,15 MHz 80 MHz
6 V in ISM bands between 0,15 MHz and 80
MHz
80% AM at 1 kHz