3i Newborn Hearing Screener IFU 024913 RevD Revision Date 08/05/2020
50
NOTICE: Deviations from the collateral standard were performed per IEC 60601-2-40
NOTICE: All necessary instruction for maintaining compliance with regard to EMC can be found in the general
maintenance section in this instruction. No further steps are required.
To ensure compliance with the EMC requirements as specified in IEC 60601-1-2, it is essential
to use only the following accessories:
ITEM
MODEL
MANUFACTURER
Cable length
[mm]
EEG Preamplifier Cable (short)
040656
Natus Medical
812 mm
EEG Preamplifier Cable (long)
030697
Natus Medical
1,727 mm
EEG Leads (used with long EEG cable only)
040740
Natus Medical
648 mm
ATA Acoustic Transducer Cable (long)
010724
Natus Medical
2,235 mm
ATA Acoustic Transducer Cable (short)
010726
Natus Medical
813 mm
40W Medical Power Supply
900611
Inventus Power
1,829 mm
AC power cord for Power Supply (US)
004535
Interpower Corp
3,048 mm
AC power cord for Power Supply (UK)
004536
Interpower Corp
3,048 mm
AC power cord for Power Supply (EU)
004537
Interpower Corp
3,048 mm
AC power cord for Power Supply (AUS)
004538
Interpower Corp
3,048 mm
NOTICE: Conformance to the EMC requirements specified in IEC 60601-1-2 is ensured if the cable types and
cable lengths are as specified above: