Table 4 - Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Power
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385 380 –390 TETRA 400
Pulse
modulation
b)
18 Hz
1,8 0,3 27
450 430 – 470
GMRS 460,
FRS 460
FM
c)
± 5 kHz deviation
1 kHz sine
2 0,3 28
710
704 – 787 LTE Band 13, 17
Pulse
modulation
b)
217 Hz
0,2 0,3 9 745
780
810
800 – 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850, LTE
Band 5
Pulse
modulation
b)
18 Hz
2 0,3 28 870
930
1,720
1,700 –
1,990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE Band
1, 3, 4, 25;
UMTS
Pulse
modulation
b)
217 Hz
2 0,3 28 1,845
1,970
2,450
2,400 –
2,570
Bluetooth,
WLAN, 802.11
b/g/n, RFID
2450, LTE Band
7
Pulse
modulation
b)
217 Hz
2 0,3 28
5,240
5,100 –
5,800
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0,2 0,3 9
5,500
5,785
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a)
For some services, only the uplink frequencies are included.
b)
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent
actual modulation, it would be worst case.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the natus XLTEK EEG32 and is the answer not in the manual?

natus XLTEK EEG32 Specifications

General IconGeneral
Brandnatus
ModelXLTEK EEG32
CategoryMedical Equipment
LanguageEnglish

Related product manuals