XLTEK EEG32U Amplifier User and Service Manual
36
Maintenance
1. Regularly inspect the trolley to ensure that casters, bolts, equipment mounting, stand shelf fasteners are always secured
tight.
2. Regularly inspect all wires and cables for cuts and damages.
3. Regularly inspect all electrical plugs to ensure they are securely inserted into their mating receptacles.
Useful Life
The EEG32U amplifier has an expected use life of at least 7 (seven) years.
Warnings and Cautions
Only use XLTEK approved equipment on the trolley/cart. Non-approved equipment may compromise the
function and safety of the system.
Do not tilt the trolley/cart more than 10° incline as this will compromise the stability of the trolley/cart.
Electrical Input and Isolation Transformer Details
EU
Electrical input
200-240V AC, 2.24A @ 50 Hz
Isolation Transformer
Powervar ABC500-22MED
North America
Electrical input
120V AC, 3.10A @ 60 Hz
Isolation Transformer Powervar ABC300-11MED