2030434-006 Rev. E EasyOne-CS Field Service Manual Page 14
9 Hidden Keys
The “hidden key” sequences access special features that are not available in the standard menu
structure of EasyOne-CS. The following table shows what key sequences are available. The column
labeled ‘Location’ indicates in what menu / screen the described key sequence has to be used.
Caution: The key sequences should not be used in conjunction with patient testing.
Location
Key
sequence
Description
Main Menu
1-3-5
Factory defaults (same as in Configuration Menu).
Device self test error
screen
1-3-5
Ignore error and go on. Use only for troubleshooting! Do not
perform a patient test, because the results could be errorneous
Device info screen (from
Configuration Menu).
1-2-4
Show additional device information.
Device info screen
1-2-5
Hard factory reset (configuration settings are reset and database is
deleted (In case of synchronisation with EasyWare, please do not
forget to synchronize EasyOne with new database.)
Device info screen
1-2-6
FVC filter disable temporarily (until next power cycle).
Device info screen
1-2-7
Switch off BTPS correction temporarily (until next power cycle).
Device info screen
1-2-8
Switch off flow linearity correction temporarily (until next power
cycle).
Device info screen
1-3-9
Restore corrupt database: Caution use only for inconsistent
databases