9 Hidden Keys
The “hidden key” sequences access special features that are not available in the standard menu
structure of EasyOne-CS. The following table shows what key sequences are available. The column
labeled ‘Location’ indicates in what menu / screen the described key sequence has to be used.
Caution: The key sequences should not be used in conjunction with patient testing.
Factory defaults (same as in Configuration Menu).
Device self test error
screen
Ignore error and go on. Use only for troubleshooting! Do not
perform a patient test, because the results could be errorneous
Device info screen (from
Configuration Menu).
Show additional device information.
Hard factory reset (configuration settings are reset and database is
deleted (In case of synchronisation with EasyWare, please do not
forget to synchronize EasyOne with new database.)
FVC filter disable temporarily (until next power cycle).
Switch off BTPS correction temporarily (until next power cycle).
Switch off flow linearity correction temporarily (until next power
cycle).
Restore corrupt database: Caution use only for inconsistent
databases