Lithotripsy, External Defibrillation, Ultrasound Procedures and Radiation - If you are
required to undergo lithotripsy, external defibrillation, high-output ultrasound,
radiation therapy, or ultrasonic scanning, inform the medical personnel conducting
the procedure that you have an implanted SCS system and follow these precautions:
• Turn off the IPG before the procedure.
• Have medical personnel use the equipment as far away from the IPG
as possible.
• Have medical personnel keep fields, such as current, radiation, or
high-output ultrasonic beams, away from the IPG.
• Equipment should be set to the lowest energy setting possible.
• After the therapy or procedure, check to see that the IPG is
functioning properly by gradually increasing the IPG’s stimulation to
the desired level.
• If you suspect that the device is not functioning properly after the use
of these therapies or procedures, please contact your doctor.
Radio-Frequency or Microwave Ablation – An electrical current produced by a
radio/micro wave is used to heat up a small area of nerve tissue, thereby decreasing
pain signals from that specific area.
Radio-Frequency or Microwave Ablation – Safety has not been established for Radio-
Frequency or Microwave Ablation in people who have an implanted neurostimulation
system. Induced electrical currents may cause heating, especially at the lead electrode
site, resulting in tissue damage.
Patient Manual 11052 Rev A (2015-01-16) 14