Do you have a question about the NI 9211 and is the answer not in the manual?
Install NI-DAQmx 9.2+ and select application software support for calibration.
Lists manuals and help files for NI 9211, NI-DAQmx, and application software.
Details equipment like calibrators and chassis required for verification and adjustment.
Specifies required ambient temperature, humidity, and connection practices.
Guide to install NI-DAQmx, connect hardware, and perform initial checks.
Procedure to verify NI 9211 operation using specified test points.
Steps to determine the NI 9211's As-Found status via calibrator measurements.
Outlines the sequence of operations required to adjust the NI 9211.
Procedure to adjust the accuracy performance of the NI 9211.
Information on updating the NI 9211 internal calibration memory (EEPROM).
Repeat verification steps to determine the As-Left status after adjustment.
Accuracy table based on calibrated scaling coefficients stored in onboard EEPROM.
National Instruments website and contact information for technical support.
Install NI-DAQmx 9.2+ and select application software support for calibration.
Lists manuals and help files for NI 9211, NI-DAQmx, and application software.
Details equipment like calibrators and chassis required for verification and adjustment.
Specifies required ambient temperature, humidity, and connection practices.
Guide to install NI-DAQmx, connect hardware, and perform initial checks.
Procedure to verify NI 9211 operation using specified test points.
Steps to determine the NI 9211's As-Found status via calibrator measurements.
Outlines the sequence of operations required to adjust the NI 9211.
Procedure to adjust the accuracy performance of the NI 9211.
Information on updating the NI 9211 internal calibration memory (EEPROM).
Repeat verification steps to determine the As-Left status after adjustment.
Accuracy table based on calibrated scaling coefficients stored in onboard EEPROM.
National Instruments website and contact information for technical support.
The NI 9211 is a calibration procedure document that outlines the steps for verifying and adjusting the performance of the National Instruments 9211 module. This document is essential for maintaining the accuracy and reliability of the NI 9211 in various test and measurement applications.
The primary function of the NI 9211 calibration procedure is to ensure that the NI 9211 module operates within its published specifications. This involves a two-part process: performance verification and accuracy adjustment.
Performance Verification: This step determines the "As-Found" status of the NI 9211. It involves connecting the module to a high-precision calibrator and acquiring samples at specified test points. The average samples are then compared against predefined test limits to ascertain if the module is performing correctly. This verification process helps identify any deviations from the expected performance before any adjustments are made. The procedure details how to configure an AI voltage channel, set timing parameters, and acquire/average samples for each channel.
Accuracy Adjustment: If the verification process reveals that the NI 9211 is not meeting its specifications, an adjustment procedure is performed. This involves using a calibrator to provide precise reference values to the module and then calling an adjustment function within the NI-DAQmx software. The adjustment process updates the internal calibration memory (EEPROM) of the NI 9211, ensuring that its measurements are accurate. The document specifies the reference value to be used for adjustment and the physical channels involved.
The NI 9211 calibration procedure is designed for users who need to maintain the accuracy of their NI 9211 modules. Key usage features include:
The NI 9211 calibration procedure itself is a key maintenance feature, ensuring the longevity and reliability of the NI 9211 module.
| Number of Channels | 4 |
|---|---|
| Resolution | 24 bits |
| Connectivity | NI CompactDAQ, NI CompactRIO |
| Operating Temperature Range | -40 °C to 70 °C |
| ADC Resolution | 24 bits |
| Input Range | ±80 mV |
| Isolation | Channel-to-earth ground |
| Connector Type | Screw terminal |
| Storage Temperature | -40 °C to 85 °C |
| Relative Humidity | 10% to 90% RH, noncondensing |
| Input Type | Thermocouple |
| Sample Rate | 14 S/s |
| Sampling Rate | 14 S/s per channel |
| Accuracy | ±0.3 °C |
| Temperature Sensor Compatibility | J, K, T, E, N, B, R, S |
| Isolation Voltage | 60 VDC |
| Shock | 30 g |
| Vibration | 5 g |
| CJC Sensor | Integrated cold-junction compensation (CJC) sensor |