7
After the last standard value input screen, Press “MENU/ENTER”
to view the confirmation selection. Here you can select YES or NO
with the “MODE/UP” or “DOWN” keys. Press “MENU/ENTER” to
confirm the desired selection.
8.2 Sensitivity Calibration
Sensitivity Calibration is specific to higher capacity models where
the sensor has a mV/V value equal to a torque value. Input in the
values by pressing “MODE/UP” to adjust the value, touch “SAVE/
PRINT” to shift to the next position. Press “MENU/ENTER” to view
the confirmation selection. Here you can select YES or NO with
the “MODE/UP” or “DOWN” keys. Press “MENU/ENTER” to con-
firm the desired selection.
8.3 Weight Calibration
In this mode, “Weights” means mass weight. When a lever and a
calibrated mass weight are used creating a standard torque, this
mode calculates the value properly to calibrate the instrument.
The torque generated by weights on a lever (L) is associated with
the local gravity acceleration (g) that the user inputs.
Input in the standard lever length (L), gravity (g) and weight val-
ues by pressing “MODE/UP” to adjust the value, touch “SAVE/
PRINT” to shift to the next position. Press “MENU/ENTER” to view
the next available standard value input screen.
After the last standard value input screen, Press “MENU/ENTER”
to view the confirmation selection. Here you can select YES or
NO with the “MODE/UP” or “DOWN” keys. Press “MENU/ENTER”
to confirm the desired selection.
8.4 Force Calibration
When using a standard force (or standard calibrated force via a
weight) and a lever (L), select Force Calibration. The standard
force applied to the lever may be produced by force from weights,
a test machine or force gauge that applies a standard force.
Input in the standard lever length (L) and force values by pressing
“MODE/UP” to adjust the value, touch “SAVE/PRINT” to shift to
the next position. Press “MENU/ENTER” to view the next available
standard value input screen.
After the last standard value input screen, Press “MENU/ENTER”
to view the confirmation selection. Here you can select YES or NO
with the “MODE/UP” or “DOWN” keys. Press “MENU/ENTER” to
confirm the desired selection.