Why the characters' positions on the screen of Nidek Medical Equipment are shifted?
- JJason HillSep 7, 2025
If the characters' positions on the screen of your Nidek Medical Equipment are shifted, the backup battery may be dead. Contact NIDEK.
Why the characters' positions on the screen of Nidek Medical Equipment are shifted?
If the characters' positions on the screen of your Nidek Medical Equipment are shifted, the backup battery may be dead. Contact NIDEK.
What to do if Nidek Medical Equipment printing out cannot be performed?
If your Nidek Medical Equipment isn't printing, it could be due to a few reasons: * The printer paper may have run out. Check and replace the paper if necessary. * The printer paper may be jammed. Remove any jammed paper. * The measurement data might be deleted. Check the memory lamp to see if the measurement data is stored in the instrument.
Why Nidek Medical Equipment screen does not turn ON though the instrument is powered ON?
If the screen on your Nidek Medical Equipment doesn't turn on even though the instrument is powered on, it could be due to the following: * The monitor auto-OFF function may be activated. If the memory lamp is blinking, try pressing any switch. * The brightness adjustment may not be correct. Adjust the brightness using the control panel under the cover for the sub-switches.
What to do if Nidek Medical Equipment screen suddenly turns OFF?
If the screen on your Nidek Medical Equipment suddenly turns off, here's what might be happening: * The monitor auto-OFF function is activated. If the memory lamp is blinking, press any switch. * The fuses are blown. Check the fuses and replace them if necessary.
Why the main unit of Nidek Medical Equipment cannot be moved horizontally?
If the main unit of your Nidek Medical Equipment cannot be moved horizontally, it is possible that the locking knob is set to LOCK or BRAKE. Cancel the LOCK or BRAKE of the locking knob.
What to do if Nidek Medical Equipment printer is working, but nothing is printed on the paper?
If your Nidek Medical Equipment printer is working, but nothing is printed on the paper, it could be because: * The printer paper is set upside down. Reverse the printer paper. * The printer head is damaged. If reversing the paper doesn't solve the issue, contact NIDEK.
What does ERR9:CHARGE ERROR mean for Nidek Medical Equipment?
If you see 'ERR9:CHARGE ERROR' on your Nidek Medical Equipment, it may be caused by a failure of the transformer or the driver board, or break of the cable. Turn OFF the power of the instrument and contact NIDEK or your authorized distributor.
What does ERR2:BATT ERROR mean for Nidek Medical Equipment?
If you see 'ERR2:BATT ERROR' on your Nidek Medical Equipment, the battery may have run out of power, or there may be a failure of the main board or the Real Time Clock on the main board. Turn OFF the power of the instrument and contact NIDEK or your authorized distributor.
Why the printed date, time and patient No. of Nidek Medical Equipment are not correct?
If the printed date, time, and patient number are not correct on your Nidek Medical Equipment, the backup battery may be dead. Contact NIDEK.
Why the printed characters of Nidek Medical Equipment are imperfect?
If the printed characters from your Nidek Medical Equipment are imperfect, the printer head may be damaged. Contact NIDEK.
Brand | Nidek Medical |
---|---|
Model | NT-4000 |
Category | Medical Equipment |
Language | English |
Provides an overview of the NT-2000 tonometer, its components, and key features.
Explains the scientific principles and optical detection methods underlying the NT-2000's operation.
Specifies the intended medical purpose and applications for the NT-2000 device.
Details the safety, electrical, and operational classifications of the instrument.
Defines the meaning of various symbols and icons used throughout the operator's manual.
Provides essential guidelines for the safe handling, transportation, and initial setup of the instrument.
Outlines critical instructions for proper electrical wiring and safe connection of the device.
Details crucial warnings and precautions to be observed during the operation of the NT-2000.
Specifies procedures for safe practices after use and during routine maintenance checks.
Offers guidance on the environmentally responsible and legally compliant disposal of the instrument.
Describes the integrated safety mechanisms designed to protect the user and patient.
Explains the purpose and location of various warning and information labels on the instrument.
Presents a high-level flowchart illustrating the sequence of steps for basic instrument operation.
Details the step-by-step process for conducting a standard intraocular pressure measurement.
A detailed guide for performing a standard intraocular pressure measurement with the device.
Identifies common error messages and provides instructions for their resolution.
Explains how to configure instrument settings and modes using the sub-switch controls.
Instructions on setting and utilizing the eyelid detection feature for improved accuracy.
Outlines the procedure for testing and verifying the accuracy of the air pressure output.
Guide on how to adjust various operational parameters and settings of the instrument.
Provides instructions for the cleaning and maintenance of critical optical and air delivery components.
Specific steps for cleaning the air nozzle to ensure measurement accuracy.
Lists common operational problems and suggests solutions for troubleshooting.
Details specific error messages displayed by the instrument and their corresponding countermeasures.
Step-by-step instructions for replacing the printer paper roll when it runs out.
Guide on how to safely access and replace internal fuses if the instrument does not power on.
Instructions on how to properly attach the disposable paper to the patient's chinrest.
Procedures for cleaning the external surfaces of the instrument to maintain hygiene and appearance.
Presents detailed technical specifications, including measurement ranges, alignment methods, and display.
Lists all the components and accessories included in the standard package of the NT-2000.
Describes optional accessories that can be purchased to enhance the functionality of the instrument.