______________________________________________________________________________________________________________
2010-2324CE-B September 2015 Page 1 of 12
USER'S GUIDE
Mark 5 Nuvo Family
NUVO, NUVO 8 & NUVO 10
OXYGEN CONCENTRATOR
[Original language is English]
Federal Law (US) restricts this device to sale
by, or on the order of, a licensed physician.
This oxygen concentrator should be used only
under the supervision of a licensed physician.
____________________________________________________________________________________
CONTENTS GLOSSARY OF SYMBOLS
WARNING
GLOSSARY OF SYMBOLS 1
1.GENERAL SAFETY GUIDELINES 2
1.1. Method for Waste Disposal 3
1.2. Method for Disposing of Device 3
2. Description 3
2.1. Front panel (Fig. 2. 1) 3
2.2. Rear panel (Fig. 2. 2) 3
3. STARTING-UP / INSTALLATION 3
3.1. Use in direct oxygen therapy 3
4. CLEANING-MAINTENANCE 4
4.1.Cleaning 4
4.2.Everyday disinfection 4
4.3.Maintenance 4
5.USEFUL INFORMATION 4
5.1.Accessories and spare parts 4
5.2.Materials in direct/indirect contact
with patient 5
5.3.Operating principles 5
5.4.Alarms - Safety devices 5
5.5.Indicators 5
5.6.Expected Life 6
5.7.Technical characteristics 6
5.8.Standards 7
5.9.Preventative Maintenance 7
5.10 Troubleshooting 8
6. EMC, Electromagnetic statements 9
: ON (Power Switched on)
: OFF (Power Switched off)
: Type B Device
: Class II Protection
: DO NOT EXPOSE TO OPEN FIRE
: DO NOT USE OIL OR GREASE
: Technical Information
: Consult the accompanying documents
: Keep in vertical position
: Fragile - Handle with care
: Oxygen concentration warning light
0413 Complies with the 93/42/EEC directive
certified by the approved organization 0413.
DANGER: Do not smoke when using
oxygen or when near this device.
NUVO 8
s
hown for reference
.

Other manuals for Nidek Medical Nuvo

Questions and Answers

  • M
    Madison SmithAug 20, 2025
    Why does my Nidek Medical Equipment compressor stop and start?
    • S
      swansonmichaelAug 20, 2025
      This issue occurs because the compressor thermal safety device has been activated, possibly due to dirty filters or a malfunctioning fan. To resolve this, stop the device and allow it to cool down. Clean the cabinet filter and then restart the device. If the problem persists, contact your equipment supplier.
  • D
    Diane DanielsAug 22, 2025
    What to do if Nidek Medical Equipment I-O button is on but there is no flow and alarm sounds?
    • K
      Keith HartmanAug 22, 2025
      If the I-O button is in the “I” (ON) position, but there is no flow and the audible alarm is sounding continuously, it indicates a broken pneumatic connection or another pressure problem. Stop the device by pressing the I-O (ON/OFF) button and contact your equipment supplier.
  • T
    Teresa LopezAug 25, 2025
    What to do if Nidek Medical Equipment I-O button is on, compressor is operating, there is flow, but alarm sounds?
    • T
      Taylor DukeAug 25, 2025
      If the I-O button is in the “I” (ON) position, the compressor is operating, and there is a flow, but the audible alarm continues to sound, this could be due to an internal electrical fault, a pneumatic circuit fault, or low purity. Stop the device and contact your equipment supplier.
  • D
    Dwayne RobinsonAug 26, 2025
    How to fix interrupted oxygen flow from Nidek Medical Equipment?
    • M
      michellegarzaAug 27, 2025
      An interruption in the oxygen enriched air flow at the nasal cannula outlet may be due to a disconnected tube or a loose humidifier cap. Ensure that all tubing connections are secure and that the humidifier is properly sealed.
  • W
    Wendy BaileyAug 27, 2025
    What to do if Nidek Medical Equipment I-O button is on but device does not operate?
    • L
      Lindsey HernandezAug 28, 2025
      If the I-O (ON/OFF) button is in the “I” (ON) position but the device does not operate, it might be due to the power cable not being correctly plugged into the wall outlet. Check the cable connection to ensure it is properly connected.
  • A
    ayaladonnaAug 29, 2025
    How to troubleshoot irregular flow at nasal cannula outlet of Nidek Medical Equipment?
    • B
      Brenda SanchezAug 29, 2025
      Irregular flow at the nasal cannula outlet is often caused by kinked or restricted cannula tubing. Straighten the tubing. If the tubing is damaged, contact your equipment supplier.
  • C
    calhoundanielleAug 31, 2025
    What to do if Nidek Medical Equipment I-O button is on, compressor is operating, there is flow, but green light is not lighted?
    • T
      Taylor ChavezSep 1, 2025
      If the I-O button is in the “I” (ON) position, the compressor is operating, and there is a flow, but the green light is not lit, it suggests a faulty indicator. Contact your equipment supplier for assistance.
  • H
    Heidi ChapmanSep 2, 2025
    Why Nidek Medical Equipment capacitor is not charged?
    • C
      Carla TaylorSep 2, 2025
      If the capacitor is not charging, it could be due to an internal electrical fault. Check the circuit breaker on the front of the unit and reset it if necessary. If the issue persists, contact your equipment supplier.
  • M
    Michelle CarterSep 4, 2025
    What to do if alarm test does not work on Nidek Medical Equipment?
    • A
      Alicia KelleySep 4, 2025
      If the alarm test is not working, it may be due to the capacitor not being charged, possibly caused by an internal electrical fault. Plug the unit in for 10 minutes and retest. If it still fails, contact your equipment supplier.
  • J
    Jacqueline McgrathSep 6, 2025
    What to do if the oxygen enriched air flow is interrupted at the nasal cannula outlet in my Nidek Medical Oxygen Equipment?
    • W
      wilsonelizabethSep 6, 2025
      If the oxygen enriched air flow is interrupted at the nasal cannula outlet of your Nidek Medical Oxygen Equipment, check that tubing connections are secure and that the humidifier is sealed.