May 2011 12010-8400 Rev F
USER'S GUIDE
Mark 5 Nuvo (Std)
LITE
OXYGEN CONCENTRATOR
Federal Law (US) restricts this device to
sale or use by, or on the order of, a licensed
physician. This oxygen concentrator should
be used only under the supervision of a
licensed physician.
IV. USEFUL INFORMATION...................................4
IV.1. Accessories and spare parts.......................................4
IV.2. Materials in direct or indirect contact with patient....4
IV.3. Operating principles..................................................4
IV.4. Alarms - Safety devices............................................5
IV.5. Indicator light function ...........................................5
IV.6. Technical characteristics...........................................5
IV.7. Standards..................................................................6
IV.8. Method for disposing of waste.................................6
IV.9. Method for disposing of device................................6
IV.10. Troubleshooting........................................................7
APPENDIX A EMC INFORMATION......................8,9
CONTENTS
: ON (power switched on)
: Off (power switched off).
: Type B device
: Class II protection
: Do not expose to open  ames.
: Do not use oil or grease.
: Technical information.
: Consult the accompanying documents.
: Keep in the vertical position.
: Fragile - handle with care.
GLOSSARY OF SYMBOLS
[Original language is English ]
GLOSSARY OF SYMBOLS..........................................1
GENERAL SAFETY GUIDELINES............................2
I. DESCRIPTION.....................................................2
I.1. Front panel (Fig. I. 1)
............................................ 3
I.2. Rear panel (Fig. I. 2)................................................3
II. STARTING-UP / INSTALLATION .....................3
II.1. Use in direct oxygen therapy...................................3
III. CLEANING-MAINTENANCE ............................4
III.1. Cleaning...................................................................4
III.2.
Everyday disinfection...................................................4
III.3. Maintenance.............................................................4
Danger: Do not smoke when using oxygen
or when near this device.
StockCheck.com

Other manuals for Nidek Medical Mark 5 Nuvo Lite

Questions and Answers

  • T
    thernandezAug 15, 2025
    What to do if Nidek Medical Equipment does not operate with the I-O button ON?
    • L
      laurapaceAug 15, 2025
      If the Nidek Medical Equipment does not operate even when the I-O button is in the ON position and an intermittent alarm sounds, it may be due to the power cable not being correctly plugged in or a power failure. Check the cable connection and the circuit breaker on the front of the unit.
  • R
    Ronnie RichardsAug 18, 2025
    Why does the compressor on my Nidek Medical Equipment stop and start?
    • B
      barnesjenniferAug 18, 2025
      If the compressor of your Nidek Medical Equipment stops in mid-cycle and restarts after a few minutes, it may be because the compressor thermal safety device has been activated, the fan is not working, or the filters are dirty. Stop the device and allow it to cool down. Clean the cabinet filter and try starting it again. Reset the circuit breaker. If it still doesn't start, it is advised to contact your equipment supplier.
  • D
    Debbie OrtegaAug 21, 2025
    What to do if there is no flow with Nidek Medical Equipment and the alarm sounds?
    • J
      Jason WrightAug 21, 2025
      If the I-O button is ON, the compressor is operating, but there is no flow and the audible alarm sounds continuously on your Nidek Medical Equipment, it could be due to a broken pneumatic connection or another pressure problem. Stop the device by pressing the I-O (ON/OFF) button and contact your equipment supplier.
  • B
    bhallAug 25, 2025
    How to fix interrupted oxygen flow in Nidek Medical Equipment?
    • C
      catherinegloverAug 25, 2025
      If the oxygen enriched product flow is interrupted at the nasal cannula outlet of your Nidek Medical Equipment, it might be due to a disconnected tube or a humidifier that isn't tight. Ensure that all tubing connections are secure and that the tubing is not kinked.
  • E
    elliottarielAug 27, 2025
    Why does the audible alarm sound continuously on Nidek Medical Equipment?
    • H
      heatherjohnsonAug 27, 2025
      If the I-O button is ON, the compressor is operating, there is a flow, but the audible alarm sounds continuously on your Nidek Medical Equipment, it could be due to an internal electrical fault or a pneumatic circuit fault. Stop the device and contact your equipment supplier.
  • J
    jennifer97Aug 30, 2025
    Why is the flow irregular from my Nidek Medical Equipment cannula?
    • J
      Joshua AndersonAug 30, 2025
      If the flow at the nasal cannula outlet of your Nidek Medical Equipment is irregular, it could be because the cannula tubing is kinked. Straighten the tubing; if it's damaged, contact your equipment supplier.
  • J
    joseph35Sep 2, 2025
    What does it mean when the red light stays on Nidek Medical Equipment?
    • R
      Rodney CherrySep 2, 2025
      If the red light remains lighted on your Nidek Medical Equipment, the product pressure may be too low or too high. Contact your equipment supplier.
  • J
    Joseph MillerSep 5, 2025
    Why doesn't the alarm test work on my Nidek Medical Equipment?
    • J
      James RiosSep 6, 2025
      If the alarm test does not work on your Nidek Medical Equipment, it may be because the capacitor is not charged or there is an internal electrical fault. If the backup capacitor has discharged, operate the unit for approximately 10 minutes and retest. Otherwise, contact your equipment supplier.
  • C
    Courtney JensenSep 8, 2025
    Why is the green light off on my Nidek Medical Equipment?
    • J
      Jaime FordSep 8, 2025
      If the compressor operates and the I-O button is in the ON position, but the green light is not lighted on your Nidek Medical Equipment, it could be a faulty indicator. Contact your equipment supplier.
  • B
    Brandy SalazarSep 12, 2025
    What does a rapidly pulsing alarm mean on Nidek Medical Equipment?
    • N
      Nicole MaciasSep 12, 2025
      If your Nidek Medical Equipment emits a rapidly pulsing alarm, it could be related to power quality. If the problem persists, contact your power provider.

Summary

General Safety Guidelines

Use of Oxygen Safety

Safety precautions related to oxygen use, flammability, and atmospheric conditions.

Device Use and Maintenance Safety

Safety guidelines for operating and maintaining the device, including electrical safety.

Device Description and Setup

Front and Rear Panel Overview

Identification and description of components on the front and rear panels of the device.

Starting Up and Installation

Instructions for initial setup and installation of the oxygen concentrator.

Use in Direct Oxygen Therapy

Specific instructions for using the device for direct oxygen therapy.

Cleaning, Maintenance, and Useful Information

Cleaning and Disinfection Procedures

Detailed steps for cleaning the device and disinfecting accessories.

Routine Maintenance Schedule

Information on periodic maintenance tasks and recommended service intervals.

Accessories and Spare Parts

Details on compatible accessories and necessary spare parts for the device.

Materials in Patient Contact

List of materials used in parts that directly or indirectly contact the patient.

Operating Principle Explained

Explanation of how the oxygen concentrator functions to produce oxygen.

Alarms and Safety Devices

Details on the various alarms and built-in safety mechanisms of the device.

Indicator Light Functions

Explanation of the meaning of the green and red indicator lights on the device.

Technical Characteristics

Specifications such as dimensions, weight, noise level, and flow rates.

Device Specifications, Standards, and Disposal

Electrical and Environmental Specifications

Details on power supply, filters, air circulation, and environmental operating limits.

Applicable Standards and Compliance

List of standards and compliance requirements the device adheres to.

Waste and Device Disposal Methods

Guidelines for the proper disposal of device waste and the unit itself.

Preventive Maintenance Schedule

Recommended schedule for preventive maintenance tasks.

Troubleshooting and EMC Information

Troubleshooting Common Issues

Guide to identifying and resolving common operational problems with the device.

Maintenance Item Log

Information on recording maintenance activities for the device.

Electromagnetic Compatibility (EMC) Information

Guidelines and declarations regarding electromagnetic emissions and immunity.

Recommended RF Separation Distances

Guidance on safe separation distances from RF communication equipment.

Nidek Medical Mark 5 Nuvo Lite Specifications

General IconGeneral
BrandNidek Medical
ModelMark 5 Nuvo Lite
CategoryMedical Equipment
LanguageEnglish