42
Model for 220, 230, 240 V AC Areas
Input voltage: Select from 220 V, 230 V or 240 V AC by relocating the
fuse holder in the AC inlet.
Frequency: 50
-
60 Hz
Voltage fluctuation: ±10%
Rated current: 0.3 A
Fuse rating: 250 V, 0.5A, time-lag low-breaking type, 5 x 20
miniature fuse x 2
Power Cord: Use only the following power supply cord. Using the
wrong power cord could result in danger or fire. The
protection Class I equipment should be connected to
PE (protective earth) terminal.
Approved according to EU/EN standards, 3
conductor grounding Type H05VV-F, 3 m long
maximum, rated at 250 V AC minimum.
Protection Class: Class I
(7) Operating Environment
Room Temperature: 0˚ to 40˚C
Relative Humidity: 85% Max. No condensation should form.
Altitude: 2000m Max.
Pollution: Degree 2
Installation Category
(Overvoltage Category): Category 2
For indoor use only.
(8) Storage Environment
Room Temperature: -20˚ to 60˚C
Relative Humidity: 90% Max. No condensation should form.
(9) Conforming Standards
The model for 100, 110, and 120 V areas is a UL-listed product.
The model for 220, 230, and 240 V areas satisfies the EU Low-
Voltage Directive.
The model for 220, 230, and 240 V areas satisfies the EU EMC
Directive.
(The model for 100, 110, and 120 V areas is not covered by the FCC.)