V.
TROUBLE
SHOOTING
CHART
If
any
trouble
cited
below
occurs,
take
appropriate
action,
referring
to
the
following
chart,
before
requesting
our
repair
service.
1.
Optical
system
Failures
Causes-
Actions
Cutting
off
or
un¬
even
brightness
of
viewfield
Revolving
nosepiece
is
not
correctly
set;
)
objective
is
out
of
optical
path
,
Condenser
is
not
raised
to
its
limit-
Lamp
bulb
is
not
correctly
inserted-
,
Lens
surfaces
of
condenser,
objective,—
I
and/or
eyepiece
are
not
clean
Surface
of
lamp
bulb
is
not
clean-
Reset
the
nosepiece
(with
click)
Raise
the
condenser
Insert
the
lamp
bulb
correctly
Clean
the
lens
surfaces
Clean
the
bulb
surface
Oirt
is
seen
in
the
viewfield
Dirt
or
dust
on
the
lens
surfaces
of
con¬
denser.
objective
and/or
eyepiece
Dirt
on
the
specimen-
Clean
the
lens
surface
Remove
the
dirt
Poor
image
(too
low
contrast
and/or
re¬
solving
power)
No
coverglass
is
used
on
the
specimen-
Immersion
oil
is
smeared
on
the
objective
of
a
dry-system
(40X
objective
is
liable
to
accept
oil)
No
immersion
oil
is
applied
to
the
end
of-
I
10OX
objective
Air
bubbles
enter
the
oil
layer-
Too
thick
specimen
is
observed-
Too
thick
coverglass
is
used-
Unspecified
immersion
oil
is
used"
Too
large
opening
of
condenser
diaphragm
Place
a
coverglass
Clean
off
the
oil
Apply
immersion
oil
Remove
the
air
bubbles
Use
a
thinner
specimen
Use
a
coverglass
of
standard
thickness
0.17
mm
Apply
Nikon
Immer¬
sion
oil
Close
the
diaphragm
properly
I
Image
shows
poor
Condenser
diaphragm
is
excessively
closed-
details
for
too
much
contrast
Open
the
diaphragm
properly
12