EasyManuals Logo

Nippon AQUA User Manual

Nippon AQUA
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #10 background imageLoading...
Page #10 background image
Instruction Manual AQUA kit for MA-3 Solo NIC-TD-0000187-01
Nippon Instruments Corporation 8
(4) Attach the mercury collector tube to the mercury collector tube connecting tube.
(5) Drip “20 mL of distilled water and standard mercury liquidor “20mL of unknown
sample” into the impinger.
(6) Add 1 mL of 1:1 sulfuric acid and 1 mL of 10% stannous chloride solution.
(7) Select “COLLECTOR TUBE / AQUA kitas the measurement method. Click START
button.
(8) Close the inside cylinder of the impinger.
* Check if it is bubbling.
If not, there is a leakage. Open the impinger inside cylinder, stop the sampling, and
fix the leakage. (Refer to 5.4 LEAK CHECK)
(9) When the measurement is completed, remove the inside cylinder of the impinger or
remove Mercury Collector Tube.
(10) Attach the mercury collector tube to the holder. Measure it with MA-3 Solo.
(For the detail, refer to the instruction manual of MA-3 Solo)
METHOD: COLLECTOR TUBE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nippon AQUA and is the answer not in the manual?

Nippon AQUA Specifications

General IconGeneral
BrandNippon
ModelAQUA
CategoryMeasuring Instruments
LanguageEnglish