Instruction Manual AQUA kit for MA-3 Solo NIC-TD-0000187-01
Nippon Instruments Corporation 9
The calibration curve can be created by measuring the standard solution by the thermal
vaporization method or mercury collector tube measurement by using this unit.
However, the atomization method can be varied due to the deterioration or the filling
condition of the mercury collector tube. For the accurate measurement, create the
calibration curve by mercury collector tube measurement by using this unit.
(11) Discard the effluent after the measurement into a tank or the like, and clean the outside
cylinder and inside cylinder of the impinger.
* If the stannous chloride solution is left, mercury will volatilize during the next
measurement and a precise and accurate measurement will be impossible.
<Sampling of 2
nd
time and after>
Repeat step 4) - 11).
<To perform the sampling and measurement at the same time>
1) Install the mercury collector tube to measure to MA-3 Solo.
2) Perform the sampling.
3) After the measurement, remove the mercury collector tube from MA-3 Solo, and
install the mercury collector tube that has finished sampling.