NIPRO Diagnostics True2go - Obtaining a Blood Sample

NIPRO Diagnostics True2go
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6. Wait until Drop Symbol appears in
Display. Keep Strip in Meter until
testing is finished.
Note: If Strip is removed before testing
is finished, an error message
appears. Release and discard old
Strip. Use new Strip for testing.
7. Turn Control bottle upside down.
Squeeze one drop of Control onto
a clean tissue. Wipe off bottle tip.
8.
Gently squeeze a Control drop onto
a small piece of unused aluminum
foil or clear plastic wrap. Dispose
after use.
9.
With Strip still in Meter, touch edge
of Sample Tip to drop of Control.
Allow drop to be drawn into Strip.
10. Remove Strip from drop. D
ashes
appear across the Display to show
Meter is testing.
Note: If Meter does not begin testing
soon after drawing up sample,
release and discard Strip. Repeat
test with new Strip. If problem
persists, see Troubleshooting.
18
LB0010
9 JUNE 2009
Meter Testing
Drop Symbol
T2g_OB_GutsENG_F4NPD04r21.indd 18 12/21/11 9:56 AM