24
System Out of Range Warning Messages
WARNING!
Meter reads blood glucose levels from
20-600 mg/dL.
If blood test result is less than 20 mg/dL,
Lo” appears in Meter Display.
If blood test result is greater than 600
mg/dL, “HI” appears in Meter Display.
ALWAYS repeat test to confirm Low (“Lo”) and
High (“HI”) results. If results still display “Lo
or “HI”, call your Doctor or Diabetes Healthcare
Professional immediately.
Note:Lo” results are included in the Average
as 20 mg/dL. “HI” results are included as
600 mg/dL.