Health Related Information
For Healthcare Professionals:
If comparing results between TRUEtrack and a lab
system, perform a TRUEtrack blood test within 30
minutes of lab test. Results from TRUEtrack are
considered accurate if within +/– 20% of lab results.
5
If patient has recently eaten, fingerstick results from
TRUEtrack can be up to 70 mg/dL higher than
venous lab results.
6
Do not use venous blood with
TRUEtrack.
Instructions for use of the System for more
than one person may be found in the Quality
Assurance/Quality Control Manual. The QA/QC
Manual is available, call 1-800-803-6025 or
1-954-677-4599.
6