PULSE 2.0 User Manual
11
Atmospheric pressure (storage and transportation): 190hPa to 1060hPa
Atmospheric pressure (operating): 700hPa to 1060hPa
AC-DC ADAPTER
WARNING! Only use the AC-DC adapter model number 30120 provided with the system. Using a dierent adapter may
cause the system to not operate correctly.
Input: 100-240V
0.8-0.4 A 50/60 Hz per NormaTec model number 30120
Output: 15V DC minimum 1.6 A per NormaTec model number 30120
ELECTROMAGNETIC COMPATIBILITY >>
The information contained in this section (such as separation distances) is in general specifically written with regard to the
NormaTec PULSE 2.0. The numbers provided will not guarantee faultless operation but should provide reasonable assurance
of such. This information may not be applicable to other medical electrical equipment; older equipment may be particularly
susceptible to interference.
GENERAL NOTES
Medical electrical equipment requires special precautions regarding electromagnetic compatibility (EMC) and needs to
be installed and put into service according to the EMC information provided in this document and the remainder of the
instructions for use of this device.
Portable and mobile RF communications equipment can aect medical electrical equipment.
Cables and accessories not specified within the instructions for use are not authorized. Using other cables and/or accessories
may adversely impact safety, performance, and electromagnetic compatibility (increased emission and decreased immunity).
Care should be taken if the equipment is used adjacent to or stacked with other equipment; if adjacent or stacked use is
inevitable, the equipment should be observed to verify normal operation in the configuration in which it will be used.
ELECTROMAGNETIC EMISSIONS
This equipment is intended for use in the electromagnetic environment specified below. The user of this equipment should
assure that it is used in such an environment.
EMISSIONS
COMPLIANCE
ACCORDING TO
ELECTROMAGNETIC ENVIRONMENT
RF emissions
(CISPR 11)
Group 1 The equipment uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
CISPR emissions classification Class B The equipment is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
(IEC 61000-3-2)
Class A
Voltage fluctuations/flicker
(IEC 61000-3-3)
Complies

Other manuals for NormaTec PULSE 2.0

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NormaTec PULSE 2.0 and is the answer not in the manual?

NormaTec PULSE 2.0 Specifications

General IconGeneral
Compression Levels7
Bluetooth ConnectivityYes
Zone Boost FeatureYes
MaterialNylon
PowerRechargeable Lithium-Ion Battery
Pressure Range30-110 mmHg
Product NameNormaTec PULSE 2.0
CategoryPersonal Care Products
TypeRecovery System
TechnologyPneumatic Compression
Zones5
Charging Time3 hours
CompatibilityiOS and Android
Mobile AppYes
Warranty1 year