PULSE User Manual
15OPERATING INSTRUCTIONS
TO STOP OR PAUSE THE SESSION
To stop the session at any time, press the Play/Pause button. This will pause the session. If you are
done using the system, remove the Attachments from Junction Box, remove the Attachment(s) from
your limb(s), turn o the device, and disconnect the hose from PULSE Control Unit.
To restart your paused session, press the Play/Pause Button again.
To disconnect connectors from the Junction Box or the PULSE Control Unit, push on the button
on the top of the NormaTec Hose Connector/ Attachment Connectors while pulling away from the
Junction Box/PULSE Control Unit.
TO TURN OFF THE DEVICE
To turn o the system, press the Power Button and confirm that the green power LED is shut o.
Unplug the power supply from the system and the wall.
HOW TO USE ZONE BOOST
During your session, you can increase the intensity of a single zone with the Zone Boost feature.
Zone Boost will add an extra 60 seconds of massage time, as well as 10 mmHg increased pressure,
in the zone that is active when the Zone Boost Button is pressed. Zone Boost can be activated when
a session is running, can only be active in one zone at a time, and is available with the NormaTec
patented Pulse massage pattern.
To determine which zone is active, refer to the graphic on the display screen—the active zone
will be flashing.
Step 1: When the massage is active in the zone you want to boost, press the Zone Boost button. The
white Zone Boost indicator will appear in the center of the active zone and the message “Zone Boost
Active” will appear below the intensity level on the screen.
When the massage moves to the next zone in the Attachment, the white Zone Boost indicator
will remain solid in the boosted zone, however, zone boost will not be active until the massage
pattern returns to the selected zone.
Step 2: To turn o Zone Boost at any time, press the Zone Boost button. The white Zone Boost
indicator and the “Zone Boost Active” message below the intensity level will disappear from the
screen.
When you power o the machine, the Zone Boost feature will be turned o.
USING THE INTERNAL BATTERY
The battery is designed to allow use of the NormaTec PULSE without being next to a power outlet.
While great for day trips, the battery is not designed to store a charge for long periods of time and will
discharge from 100% to 0% over 24 hours. It takes approximately six hours to fully charge the battery
when the device is plugged in and not in use. A fully charged battery will run the unit for up to 2 hours.