4.1.3 Symbols
Instead of the current measurement information in yellow colour, some symbols
might be displayed. Please double-check, which chamber is affected before you
try to solve the problem so that you do not stop an on-going measurement in
another chamber accidentally.
Those symbols have the following meaning:
4.2
Main Menu Screen
The main menu screen provides easy access to programming and measurement
functions. If additional, external chambers (LabClick-aw) are connected to the
instrument, each individual one can be programed but a chamber selection has
to be done
Chamber
selection
Switch OFF
the instrument
Chamber
Status Settings
Start
functions
Switch
to measurement
screen
Symbol Description Remarks
The system detected a problem with
the respective chamber
Press on the symbol to gather more specifi c
information
Sensor will be put into operating condi-
tions
• Press on the symbol to gather more
specifi c information
• A measurement can be started even
if the symbol is present but data will
just be analyzed once the symbol has
disappeared
Based on a specifi c situation, no mea-
surement values can be established
(i.e. activated dew protection of the
sensor)
• Press on the symbol to gather more
specifi c information
• A dew protection can as well be activa-
ted, if the sensor has been calibrated
improperly and measures a much high-
er aw-value than the real one.
Danger of water condensation in the
chamber
• Remove the sample as soon as possi-
ble from the chamber
• In addition, there is a warning beep
which lasts till the opening of the cham-
ber or for 30 sec.