18
5-3 Washing and disinfecting
The Motor can be washed via Thermo Disinfector.
If a Thermo-Disinfector is not available, proceed to “5-4 Sterilization” below.
The following conditions are recommended for Thermo Disinfector: (It is recommended to use a thermo disinfector compatible with ISO 15883)
a. Cleaning: at 45˚C for 3 minutes
b. Intermediate rinse: At least 1 minute (using distilled, deionized or demineralized water)
c. Thermal disinfecting: at 93˚C for 5 minutes (using distilled, deionized or demineralized water)
d. Drying: at least 10 minutes
Prior to use of a Thermo Disinfector, read the Thermo Disinfector Operation Manual carefully.
Do not use a chlorinated cleaner.
Do not clean the Motor by an ultrasonic cleaning apparatus.
After cleaning and prior to sterilization, check the Motor in the same method as in Step “3. Check before Operation”.
The P200-WPD is made out of aluminum, therefore take special precaution to always use a PH neutral detergent compatible with
aluminum, when washing it in a Thermo Disinfector.
CAUTION