5-1 Preparation prior to washing
1) Remove the Attachment/Handpiece from each Motor.
2) Disconnect the Motor Cord from the Control Unit.
3) Check for any damage or deterioration on the outer surface of each Motor or Motor Cord, and for any deformation or loss of each connecter pin.
5-2 Cleaning
1) Wipe stains off the surface of the Motor (including the Motor Cord) thoroughly using a lint-free cloth moistened with neutral detergent for medical use or
a non-metallic brush.
2) Remove the detergent solution using a lint-free cloth moistened with distilled, deionized or demineralized water.
Perform cleansing and sterilization after every use.
Be sure to clean and sterilize the Motor after use. Unremoved blood may result in product failure due to the internal blood coagulation.
In order to prevent infection, put on a mask, rubber gloves and a pair of protection goggles.