Guidance and manufacturer's declaration - electromagnetic immunity
The Primado2 is intended for use in the electromagnetic environment specified below. The customer or the user of the Primado2 should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
EN/IEC61000-4-2
Electrical fast transient/burst
EN/IEC61000-4-4
Surge
EN/IEC61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
EN/IEC61000-4-11
Power frequency (50/60Hz)
magnetic field
EN/IEC61000-4-8
IEC60601 test level
±6kV contact
±8kV air
±2kV for power supply lines
±1kV for input/output
±1kV line (s) to line (s)
±2kV line (s) to earth
<5% Ut (>95% dip in Ut) for
0.5 cycle
40% Ut (60% dip in Ut) for 5
cycles
70% Ut (30% dip in Ut) for 25
cycles
<5% Ut (>95% dip in Ut) for 5
secs
3A/m
Compliance level
±6kV contact
±8kV air
±2kV for power supply lines
±1kV for input/output
±1kV line (s) to line (s)
±2kV line (s) to earth
<5% Ut (>95% dip in Ut) for
0.5 cycle
40% Ut (60% dip in Ut) for 5
cycles
70% Ut (30% dip in Ut) for 25
cycles
<5% Ut (>95% dip in Ut) for 5
secs
3A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of
the Primado2 requires continued operation during power mains interruptions, it is recommended
that the Primado2 be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical
commercial or hospital environment.
NOTE ‘Ut’ is the AC mains voltage prior to application of the test level