15
EN DE FR ES IT PT
WARNING
•Ifbloodinfiltratesinsideahandpiece,anautomatichandpiececleaningandlubricationsystemmaynottotally
cleantheinternalhandpiececomponentsandmayleadtobloodcoagulationinsidethehandpiece.Coagulated
bloodmaycausehandpiecefailureandoverheating,causingburninjuries.Toavoidtherisk,NSKrecommends
toimmediatelylubricatethosehandpieceswithPANASPRAYPlus/PANASPRAY.
7-6 Sterilization
Sterilize the product by autoclave sterilization. After each patient sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20min. at 121
°C
, or 15min. at 132
°C
, or 3min. at 134
°C
.
3) The handpiece should remain in the autoclave pouch until required for use.
NSK automatic handpiece cleaning and lubrication system
When using NSK automatic handpiece cleaning and lub
rication system, refer to the system instructions.