7-3 Cleaning of Optic Illumination Points
Wipe clean all Optic Illumination Points (Glass Rod) using an alcohol-immersed cotton swab.
Remove all debris and oil.
CAUTION
Do not use a sharp tool to clean the Glass Rod. It could damage the glass and reduce the light transmission.
Glass Rod
Cotton Swab
Fig. 6
7-4 Disinfection
Manual Disinfection (External Surface)
Wipe the surface with cloth soaked in disinfectant solution such as alcohol.
Automatic Disinfection (External and Internal Surfaces)
Use a thermo-disinfector in accordance with ISO 15883 (EN ISO 15883).
Refer to the thermo-disinfector manual.