10
Automatic Cleaning (External and Internal Surfaces)
Use a thermo-disinfector in accordance with ISO 15883 (EN ISO 15883).
Refer to the thermo-disinfector manual.
CAUTION
After washing with thermo-disinfector, prior to lubrication, dry the product until all internal moisture is completely removed. Thermo-disinfector
moisture remaining inside the product could reduce the effect of lubrication and could cause corrosion inside of the product.
7-2 Cleaning
Manual Cleaning (External Surface)
1) Remove debris on the handpiece surface using a brush under running water (at 38
°C
or less).
DO NOT use a wire brush.
2) W
ipe off moisture on the surface using an absorbent cloth.
CAUTION
If water enters the handpiece, using an air syringe, blow air into the rear of the handpiece to remove the internal water.