LabGard® ES Energy Saver Class II,
Type B2 Laminar Flow
Biosafety Cabinet
Models
NU-560-400/600
Bench/Console
Operation & Maintenance Manual
(Series C)
Revision 2
September 2020
Manufactured By:
NuAire, Inc.
2100 Fernbrook Lane
Plymouth, MN 55447
Toll-Free: 1-800-328-3352
In Minnesota: (763)-553-1270
Fax: (763)-553-0459
Question and Answer IconNeed help?

Do you have a question about the NuAire LabGard NU-560-600 and is the answer not in the manual?

NuAire LabGard NU-560-600 Specifications

General IconGeneral
BrandNuAire
ModelLabGard NU-560-600
CategoryLaboratory Equipment
LanguageEnglish

Summary

General Information

Safety Instructions

Outlines essential safety precautions for operating the LFBSC.

Installation Instructions

Set-Up Instructions

Covers the steps for assembling and setting up the cabinet.

Exhaust/Supply Air Checks

Procedures for verifying exhaust and supply air performance.

Certification Testing Methods and Equipment

Outlines procedures and equipment for cabinet certification.

Operating the NU-560

Biosafety Cabinet Control

Explains the BSCC system and its functions.

Standby/Run Mode Alarms

Explains the alarm system and types.

Setpoint / Limits

Configuration of airflow setpoints and alarm limits.

Calibration/Service

Accessing calibration and service functions.

Operating Guidelines

Employ Aseptic Technique

Best practices for aseptic manipulation.

Cleaning Procedures

Hazardous Drug Decontamination Procedures

Specific procedures for decontaminating after use with hazardous drugs.

General Maintenance

Decontamination

Procedures for microbiological decontamination.

HEPA Filter/Motor Replacement

Supply Filter Replacement

Step-by-step guide for replacing the supply HEPA filter.

Exhaust Filter Replacement

Instructions for replacing the exhaust HEPA filter.

Motor/Blower Assembly Removal

Procedures for removing the motor/blower assembly.

Airflow Control System Setup and Calibration

Calibration/Service Menu

Navigating the calibration and service menu.

Setpoints/Limits

Verifying or altering airflow setpoints and alarm limits.

Airflow Calibration

Procedures for calibrating airflow sensors.

Exhaust Volume Calibration without the NU-951-012 Motorized Butterfly Valve

Calibration steps for exhaust volume without a butterfly valve.

Exhaust Volume Calibration with the NU-951-012 Motorized Butterfly Valve

Calibration steps with a motorized butterfly valve.

Downflow Calibration

Calibrating the downflow sensor.

HEPA Filter Leak Test

Airflow Smoke Pattern Test

Site Installation Assessment Tests

Cleanliness Classification Test for Pharmacy Application

Main Control Board Description & Replacement

Main Control Board Replacement

Procedures for replacing the main control board.

Digital Airflow Sensor Description & Replacement

Downflow Velocity Sensor

Description and replacement of the downflow sensor.

Exhaust Pressure Sensor

Description and replacement of the exhaust pressure sensor.

Error Messages, Troubleshooting, Option-Diagnostics & Airflow Sensor Performance Verification

Error Message Troubleshooting Guide

Calibration/Service Menu

Blower/Damper Setup

Configuration for blower and damper settings.

Service

Accessing service-related data.

Alarm Setup

Configuring alarms.

Diagnostics

Accessing diagnostic tests for system components.

Airflow Sensor Performance Verification

Optional Equipment