OM0133
28
8 - CO
2
Inject Delay Time
This value specifies the time, in seconds, for an injection
of CO
2
to be measurable at the sensor. When CO
2
is injected
into the chamber, the system delays until this period has
elapsed before making a new control decision. In this manner,
diffusion induced delays do not cause the CO
2
system to
overshoot the control set point.
9 - Door Delay Time
Determines time, in seconds, which CO
2
control is inhibited to allow
chamber condition to stabilize after the inner glass door is opened. The
outer door, and perimeter heater duty cycles are increased to 100% to
guard against condensation.
10 - Air Inject Time
This value specifies the time, in seconds, for an injection of air
into the chamber.
11- Air Inject Cycle Time
This value specifies the amount of time in minutes
between each injection of air into the chamber
12 - Print Frequency Time
This parameter specifies the frequency, in minutes that lines are
to be printed on a status report. If the frequency is specified as zero,
no report will be printed.
13 - Temperature Time Out
This value determines the time, in minutes, for the temperature to
achieve setpoint. If the temperature doesn't get to within 0.2º of
setpoint within this time period, an alarm condition is declared.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the NuAire NU-5500 and is the answer not in the manual?

NuAire NU-5500 Specifications

General IconGeneral
BrandNuAire
ModelNU-5500
CategoryAccessories
LanguageEnglish

Related product manuals