800-1060 Rev: A
Device Identification
Alerts and Alarms
Please see the User Manual for complete information and instructions regarding alerts and alarms.
Alerts
Description
Low Purity
Oxygen level has dropped below 85%
Low Battery
Remaining battery run time has reached six minutes
Unauthorized Battery
A non-O2 Concepts battery is detected
Low External Power
Voltage to the device is below 11.6 volts
Alarms
Description
No Breath
No breath has been detected for 45 seconds; Device will switch to last
continuous flow setting
Low Purity
Oxygen level has dropped below 72%
Low Battery
Battery power is depleted, device will shut down
No Flow
Flow of oxygen has stopped for 45 seconds
Unauthorized Battery
A non-O2 Concepts battery is detected, and no other power source is
available
Low External Power
Voltage to the device is below 10.6 volts
No External Power
Power has been removed from the device
High External Power
Voltage to the device is above 26 volts
Over Temperature
Pump or enclosure temperature has exceeded limits
Invalid Motor Temperature
Motor temperature has exceeded limits
Invalid Box Temperature
Internal box temperature has exceeded limits
Serial number and UDI are
located on the rear label of
the device.