Device Identification
Alerts and Alarms
Please see the User Manual for complete information and instructions regarding alerts and alarms.
Oxygen level has dropped below 85%
Remaining battery run time has reached six minutes
A non-O2 Concepts battery is detected
Voltage to the device is below 11.6 volts
No breath has been detected for 45 seconds; Device will switch to last
continuous flow setting
Oxygen level has dropped below 72%
Battery power is depleted, device will shut down
Flow of oxygen has stopped for 45 seconds
A non-O2 Concepts battery is detected, and no other power source is
available
Voltage to the device is below 10.6 volts
Power has been removed from the device
Voltage to the device is above 26 volts
Pump or enclosure temperature has exceeded limits
Invalid Motor Temperature
Motor temperature has exceeded limits
Internal box temperature has exceeded limits
Serial number and UDI are
located on the rear label of
the device.