EasyManua.ls Logo

Oakton CON 6 - Normalization Temperature; Single-Point Calibration

Oakton CON 6
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instruction Manual CON 6/TDS 6
32
6.6 Normalization Temperature
You can set the meter to normalize its conductivity measurements to a standard
temperature of either 25 °C or 20 °C.
The default value is 25 °C.
1) Enter the advanced setup as described in
Section 6.1.
2) Press or key until [t.nr °C] appears on the
LCD. Press ENTER key.
3) Press or key to select either [25.0 °C] or
[20.0 °C].
4) Press ENTER key to select. The meter will take
you back to the menu, [t.nr °C].
5) Press or key to move to the next menu or
press CAL to exit to measurement mode.
6.7 Single-Point Calibration
Single-point calibration refers to calibrating one conductivity value and uses it
for the entire 5 conductivity ranges.
By selecting [no] to single-point calibration, you can perform calibration for each
conductivity range.
1) Enter the advanced setup as described in
Section 6.1.
2) Press or key until [S.P.CA] appears on the
LCD. Press ENTER key.
3) Press or key to select either [Yes] or no].
4) Press ENTER key to select. The meter will take
you back to the menu, [S.P.CA].
5) Press or key to move to the next menu or
press CAL to exit to measurement mode.
C°
C°
C°
HOLD
ENTER
HOLD
ENTER

Related product manuals